E-Beam system for continuous surface decontamination

SKAN AG

Not Published

Continuous Exterior Surface Decontamination

The E-beam tunnel guarantees the safe entry of containers with pre-sterilized syringes into the filling line isolator. This automatic E-beam system is especially suited for continuous processes with high thru-put, up to eight containers per minute.

Air flow

  • Unidirectional, vertical air flow at the entrance to the tunnel for a particle free environment before the emitter
  • Unidirectional, vertical air flow on the "sterile side" of the tunnel for a particle free environment after the emitter (Grade A, ISO 5)
  • Direct ozone extraction at the emitter and transfer to the outside air

Transport system

  • The container (tubs) are held on their edge and actively transported
  • A special transport system of high quality stainless steel, developed for cleanroom applications.
  • There is only one tub at a time in the emitter zone

Technology of the E-Beam System

  • GMP Cleanroom Class Grade A, ISO 5 
  • Double filtered (HEPA H14) in and out of the unit 
  • DEHS testing ports for "in situ" testing 
  • Uni-directional airflow over the entire "sterile" sections of the transport system 
  • Automatic leak test/mouseholes with pneumatic gaskets 
  • Controlled differential pressure between isolator and room 
  • Special stainless steel-lead-stainless steel "sandwich" construction 
  • Stainless steel AISI 316L/304, polished to 0.8/1.2 µm 
  • Easy and quick access for cleaning and maintenance 
  • Clearly defined separation of attachment to the isolator